This is the most complete calibration procedure, with the minimum number of assumptions. The first step is to measure the power detected by the receiver on the Sky, then on a warm load (the Chopper), then on a cold (liquid nitrogen) load, averaged over the channels of a specific backend. For a perfectly linear detector, the output power is proportional to the total input noise (Receiver + Load), hence the receiver temperature is simply given by


\begin{displaymath}
Trec = \frac{T_{load} * Mean\_cold - T_{cold} * Mean\_load}
{Mean\_load - Mean\_cold}
\end{displaymath} (12)

Then, one can proceed as in TREC mode.

The description above refers to what happens when OBS when you use the CAL COLD command (in either TREC or AUTO mode).