OBS\BANDPASS [Time]
Start a specific procedure to calibrate instrumental phases of the
spectral correlator. Default integration time is 5 seconds. The
spectral correlator is automatically connected to a noise source for
this purpose. Delay and phase tracking are cancelled during the
measurement.